CONTROL-1164. |
General.
The CONTROL-1164 is used with an ISO-1064CE EEG Amplifier to form a stand-alone system with analog outputs for connection to an AD Converter (universal acquisition and analysis environment). The CONTROL-1164 converts the digital EEG signals of the EEG-1064CE amplifier into analog, adds a selectable 4th order anti-aliasing filter and fully controls the functionality and modes of the EEG-1064CE Amplifier. The impedance check of the electrodes is displayed on a panel with an LED per electrode. The CONTROL-1164 is connected by a thin and flexible microphone-type cable to the ISO-1064CE amplifier. The photo at the right is of the CONTROL-1132. Only the LED-field is different for the CONTROL-1164 |
![]() |
||
|
Controller specifications.
| Control-1164 | |
| Input selection: CAL - mV - kohm - EEG | • |
mV Full scale switch: 8 mV, 4 mV, 2 mV, 1 mV, 0.5 mV, and 0.25 mV |
• |
| 4th order anti-aliasing low pass filter: 15 Hz, 35 Hz, 70 Hz and 100 Hz | • |
| High Pass fiter, time constant: 1 sec, 0.3 sec and 0.1 sec | • |
| NOTCH filter 50Hz | • |
| Selectable calibration voltage: 10 µV, 20 µV, 50 µV, 100 µV, 200 µV, 500 µV, 1000 µV and 2000 µV |
• |
| Selectable Common Mode Test Voltage: 0.05 mV, 0.1 mV, 0.2 mV, 0.5 mV, 1 mV, 2 mV, 5 mV and 10 mV |
• |
| Electrode test indication panel | • |
Electrode test level (mV): |
• |
Electrode test level (kohm): |
• |
| AUT/MAN switch for calibration signal | • |
| POLYGRAPHY switch: selects between EEG and Polygraphy for upper channels | • |
NO BOX indicator: detection of a disconnected ISO-1064CE amplifier |
• |